Upright vs Inverted Microscopes: Design and Uses

Table of Contents

What Do Upright and Inverted Microscopes Mean?

When people compare microscope types, the first fork in the road is often whether the stand is upright or inverted. The terms describe where the objective lenses sit relative to the specimen and how illumination is delivered. This choice influences everything from what samples you can observe comfortably to the illumination methods available and the kinds of accessories you can add later.

At a glance:

  • Upright microscope: Objectives sit above the specimen and look downward. For transmitted-light methods (e.g., brightfield on glass slides), the condenser is below the specimen and sends light upward.
  • Inverted microscope: Objectives sit below the specimen and look upward. For transmitted-light methods, the condenser is above the specimen and sends light downward.
Upright microscope
Attribution: Databese Center for Life Science (DBCLS). Upright microscope: Image is from Togo picture gallery maintained by Database Center for Life Science (DBCLS).

These orientations are not just cosmetic. They condition the sample geometry and working distance, the illumination options, and how easily you can handle particular specimens (e.g., culture dishes versus thin sections). They also affect ergonomics and stability (see the section on mechanics and ergonomics).

Text schematic of upright vs inverted orientation
Upright (transmitted light):
  Objective
     ↓
  Cover glass/specimen/slide
     ↑
  Condenser

Inverted (transmitted light):
  Condenser
     ↓
  Culture dish/bottom/cover glass
     ↑
  Objective

Although both designs can be outfitted for transmitted light, epi-illumination (reflected light), and fluorescence, their native strengths differ. Uprights naturally suit thin, mounted samples (histology slides, thin geological sections), while inverted stands shine with thicker or containerized samples like live cells in Petri dishes or microfluidic devices. Metallurgical versions of each place emphasis on reflected-light inspection of opaque materials.

If you are deciding between them, start with your specimen format and illumination needs; then factor in ergonomics, accessories, and budget (full decision framework here).

Optical Architecture: Objective Orientation, Tube Lenses, and Light Paths

The optical train of a modern compound microscope is designed around the objective lens. Whether upright or inverted, the stand provides a stable, aligned pathway for illumination and imaging. The key difference is which side of the specimen the objective occupies and how the condenser is positioned.

Objective orientation and the specimen plane

Objectives define the primary image quality. In both upright and inverted stands, the objective’s front focal plane must coincide with the specimen plane to form a sharp intermediate image for the tube lens. In an upright stand, the objective looks downward and approaches the coverslip from above; in an inverted stand, it approaches from below the vessel (e.g., the bottom of a culture dish). Orientation affects:

  • Front lens protection and contamination risk: In uprights, the objective faces down; accidental contact with a wet sample is a risk during focusing. In inverted stands, the objective faces up and is exposed to airborne dust or spills; many users add splash guards and cover the turret when not in use.
  • Working distance (WD) and clearance: Inverted objectives often emphasize longer WD to look through vessel bottoms. Upright objectives for slide work can prioritize higher numerical apertures with shorter WD. For details, see Objective compatibility.

Infinity-corrected optics and tube lenses

Most contemporary research-grade stands use infinity-corrected objectives that produce collimated light between the objective and tube lens. The tube lens then forms the intermediate image sent to the eyepieces and camera. This configuration allows insertion of auxiliary optics (filters, beam splitters) without altering the objective’s nominal magnification. Whether upright or inverted, the tube lens focal length is matched to the objective series.

Practical note: Infinity systems simplify adding epi-fluorescence, DIC prisms, or teaching heads because the collimated space tolerates these modules without shifting the image plane.

Transmitted-light path

In upright transmitted-light configurations, a field diaphragm, collector optics, and a condenser direct light upward through the specimen toward the objective. In inverted configurations, these condenser components live above the stage and direct light downward. In both, Köhler illumination is the standard approach for even, high-contrast illumination: field and aperture diaphragms are conjugate to the field and pupil planes, respectively, promoting uniform fields and control of angular illumination.

Köhler Illumination with the Inverted Microscope (15174751101)
Attribution: ZEISS Microscopy from Germany.

Ask your ZEISS account manager for a lab poster! You’ll find more knowledge brochures and materials on our website www.zeiss.com/microscopy

Images donated as part of a GLAM collaboration with Carl Zeiss Microscopy – please contact Andy Mabbett for details.

Reflected-light (epi) path

Epi-illumination sends light through the objective onto the specimen and returns reflected or fluorescent light through the same objective. This geometry is available for both upright and inverted stands via an epi-illuminator with a beamsplitter or dichroic mirror. The upright’s mechanical path to the objective usually leaves more physical space for interchangeable epi-modules; inverted research stands typically integrate epi-fluorescence within a central module above the objective turret.

As you read about illumination methods below, remember the underlying architecture: the illumination mode you desire must be compatible with the stand’s available condenser or epi-illuminator modules and with your objectives.

Sample Geometry, Working Distance, and Cover Glass Considerations

Sample geometry is where upright and inverted designs most clearly diverge. What sits between the objective and your specimen? How thick is it, and what optical properties does it have? Your answers here drive the choice of stand and objectives.

Thin slides vs. containerized samples

  • Upright stands excel with thin, rigid samples like standard microscope slides (approximately 1 mm thick glass) with a #1.5 coverslip (~0.17 mm) on top. The objective engages the coverslip directly, minimizing extra material in the optical path.
  • Inverted stands are ideal when the specimen resides in a vessel: Petri dishes, multi-well plates, culture flasks, microfluidic chips, or custom chambers. The objective must focus through the vessel bottom to the specimen above.
Inverted Microscope
Attribution: Zephyris at English Wikipedia. By Richard Wheeler (Zephyris) 2007.
Zeiss ID 03 Inverted microscope for tissue culture.

Working distance and clearance

Working distance (WD) is the space between the objective’s front lens and the focal plane in the specimen. Longer WD provides clearance for vessel bottoms and reduces the chance of collision but often comes with trade-offs in maximum numerical aperture and correction level. In applications with deep or tall samples, inverted long-WD objectives allow focus without compressing or contacting the specimen support.

Conversely, when you need the highest resolution on a thin specimen, upright high-NA objectives (including immersion types) engage close to the coverslip. Inverted stands can also use immersion objectives, but care is required to manage immersion media with upward-facing lenses.

Cover glass thickness and bottom thickness

Objectives are corrected for specific cover glass thicknesses or vessel bottom thicknesses. A common standard for slide covers is #1.5 (approximately 0.17 mm). Many objectives assume this value; some have a correction collar to tune for small variations. When focusing through a culture vessel, the bottom thickness may differ substantially. Options include:

  • Glass-bottom dishes or plates with a #1.5 coverslip window, letting you use objectives corrected for 0.17 mm glass.
  • Plastic-bottom vessels with thicker bases; dedicated long-WD objectives or correction-collar objectives can compensate to a degree.
  • No-cover objectives for metallurgical work (reflected light), designed to focus directly on polished surfaces without a coverslip.

Mismatch between designed and actual thickness introduces spherical aberration, degrading contrast and resolution. If you anticipate variable vessel thicknesses, consult the objective compatibility section to plan an appropriate set of lenses.

Specimen accessibility

Inverted stands provide clear access to the top of the specimen for pipetting, micromanipulation, or environmental enclosures because the objective approaches from below. Upright stands give access from below (for condensers and transmitted-light accessories) and from above for epi-modules, while the specimen sits on the stage; this is ideal for rigid sample mounting, microtomy sections, or manual scanning of thin slides.

Illumination Modes: Transmitted, Reflected, Phase, DIC, Darkfield, Fluorescence

Both upright and inverted microscopes support a broad toolset of contrast methods. The practicality of each depends on how condensers, objectives, and epi-illuminators fit around your specimen format.

Brightfield (transmitted) and Köhler illumination

For transparent specimens, brightfield with Köhler illumination provides even, adjustable illumination. Uprights place the condenser below the stage; inverteds place it above the specimen. The condenser’s aperture diaphragm controls the angular spread of rays entering the specimen and into the objective’s pupil. In practice, maintaining proper alignment and cleanliness of the condenser and field diaphragm is crucial for uniform fields. Both designs allow this when outfitted with the appropriate transmitted-light modules.

Köhler Illumination with the Upright Microscope (15177755065)
Attribution: ZEISS Microscopy from Germany.

Ask your ZEISS account manager for a lab poster! You’ll find more knowledge brochures and materials on our website www.zeiss.com/microscopy

Images donated as part of a GLAM collaboration with Carl Zeiss Microscopy – please contact Andy Mabbett for details.

Phase contrast

Phase contrast converts phase shifts in transparent specimens to intensity differences using a phase ring in the objective and a matching annulus in the condenser. It is widely used for unstained cells and protists. Upright and inverted stands both offer phase contrast; on inverted stands used with culture vessels, phase objectives are often paired with long working distances and special condensers designed to clear dish walls while still projecting the proper annulus. Alignment of the annulus and phase plate is essential in either configuration; many stands include a phase telescope or centering tools to assist.

Nomarski DIC (differential interference contrast)

DIC enhances edge contrast in transparent specimens using polarizers and birefringent prisms to create shear and interference. This method requires compatible objectives, a suitable condenser (for transmitted DIC), and analyzers; exact components depend on the stand. Uprights and inverteds both support DIC if the frame is designed for it. Practical factors:

  • Upright DIC on slides benefits from rigid, uniform coverslip geometry and standard condensers.
  • Inverted DIC is popular for live-cell imaging in dishes; specialized condensers and prisms accommodate vessel geometry.

Because DIC relies on precise polarization and shear, mixed materials in the optical path (e.g., some plastics) can alter performance. Where possible, use glass-bottom dishes or ensure your vessel material is compatible with the DIC path.

Darkfield

Darkfield excludes direct rays from entering the objective so that only scattered light from the specimen forms the image. Uprights commonly use dry or oil darkfield condensers under the stage for slide work. In inverted configurations, darkfield for transmitted light is feasible but may be constrained by the condenser’s working distance and the thickness of the vessel bottom; dedicated inverted darkfield condensers or reflected-light darkfield for opaque specimens may be more practical.

Epi-illumination for opaque samples (reflected light)

Opaque materials—polished metals, semiconductors, and many manufactured parts—require reflected-light techniques. Both upright and inverted frames can be configured as metallurgical microscopes, hosting an epi-illuminator that directs light through the objective onto the sample surface. Choosing between upright and inverted in this context is mainly about sample size and handling: heavy or large parts rest stably on the stage of an inverted metallograph, while thin coupons and wafers are easily secured on an upright stage.

Fluorescence

Fluorescence imaging uses epi-illumination via dichroic mirrors and excitation/emission filters. The stand type mainly affects sample handling and environmental control. Uprights excel in fluorescence on prepared slides; inverteds excel in live-cell fluorescence and time-lapse imaging within enclosures or incubators. In both cases, proper filter sets, stray-light control, and clean optics are crucial. Because fluorescence signals are often weak, vibration control and light-tight enclosures can improve image stability and background levels; see maintenance and stability.

Ergonomics, Stability, and Stage/Focus Mechanics

Beyond optics, the physical form of the stand affects comfort, stability, and workflow speed. While both types have fine/coarse focus and stages, their mechanical emphases differ.

Specimen loading and access

  • Upright: You place a slide or thin section on the stage and bring the objective down to focus, or move the stage up depending on the design. This is efficient for scanning many slides and for techniques like polarized light microscopy of geology thin sections. Adding immersion media is straightforward because gravity helps keep the droplet at the coverslip-objective interface.
  • Inverted: You place a dish, plate, or flask on the stage, and the objective comes up from below. This leaves the specimen surface unobstructed for pipetting, manipulation, or environmental chambers. However, immersion media must be managed carefully since the objective faces upward; spill guards and careful technique help protect the turret.
ECHO Revolve Upright
Attribution: Timmesc. The ECHO Revolve hybrid microscope in Upright mode.

Focus drive and Z mechanics

Modern stands commonly offer fine/coarse focus within a comfortable reach. Uprights often move the stage for focusing, keeping the objective turret fixed; some designs move the nosepiece instead. Inverted stands typically move the objective nosepiece upward for focusing, holding the stage fixed for stability (especially when samples are heavy). In both cases, minimization of mechanical backlash and smooth, repeatable fine focus are important for high-magnification tasks.

Stage travel and sample navigation

Uprights used for slide scanning frequently include mechanical stages with X–Y verniers, allowing methodical raster scanning across standard slide dimensions. Inverted stages are often larger and flatter to support a range of vessel footprints (35 mm dishes, multi-well plates, microfluidic chips). Precision X–Y stages for inverted stands may use inserts to center and secure different vessel types.

Ergonomics and posture

Viewing comfort depends on eyepiece height, tilt, and hand placement. Uprights may place eyepieces higher, which can be mitigated by intermediate tilting heads. Inverteds often sit lower, with eyepieces closer to bench height; this can be more comfortable for long sessions. Trinocular heads allow camera placement without disrupting viewing height. Small adjustments—eyepiece inclination, chair height, and control placement—can significantly reduce fatigue.

Stability and environmental control

For live-cell imaging, inverted stands commonly pair with environmental enclosures and stage-top incubators that control temperature, gas composition, and humidity. This adds mass and sometimes moving air near the optical path. Strong, vibration-damped benches and careful cable management help maintain stability. Uprights can also use environmental chambers (e.g., for temperature), but the geometry tends to suit prepared slides and fixed samples, where environmental fluctuations are less critical.

Where Each Design Excels: Common Use Cases and Context

While both designs are versatile, certain scenarios naturally fit one better than the other. The following examples are descriptive and educational rather than procedural guidance.

Upright strengths

  • Prepared slide analysis: Histological sections, botanical thin sections, diatom slides, and stained cytology smears. The coverslip geometry is predictable, making brightfield, phase, and DIC straightforward.
  • Geological and materials thin sections: Polarized light microscopy of minerals typically uses a specialized upright polarizing microscope with rotating stage and analyzers.
  • Education and surveying: Scanning and annotating multiple slides in succession benefits from upright stages and slide holders.
  • High-magnification immersion work: Oil or water immersion on standard coverslips is direct in upright geometry.

Inverted strengths

  • Live-cell observation in vessels: Viewing adherent cells on the bottom of dishes or wells is what inverted stands were designed to do. The unobstructed top surface allows gentle media exchange and manipulation.
  • Microfluidics and organ-on-chip devices: Chips mounted on the stage can be plumbed and perfused from above while imaging from below.
  • Large or heavy specimens: Inverted metallurgical stands support and stabilize heavy parts for reflected-light inspection. The specimen can rest directly on the stage plate.
  • Time-lapse fluorescence: Environmental enclosures fit well around inverted frames, enabling long-duration imaging with controlled temperature and gases.

In practice, many labs use both types: an upright for slide-based methods and an inverted for vessel-based or live imaging. If you must choose one, map your primary sample formats and planned modalities to the features described in the decision framework.

Objective Compatibility: Working Distance, Cover Glass Correction, and Immersion Media

The objective lineup you can use comfortably is a major factor in choosing a stand. Compatibility encompasses mechanical fit, optical correction, working distance, and immersion methods.

Working distance and long-WD objectives

Long working distance (LWD) and extra-long working distance (ELWD) objectives provide additional clearance between the front lens and specimen plane. They are common on inverted stands to look through vessel bottoms and on metallurgical stands for reflected-light inspection of uneven surfaces. The trade-offs typically include lower maximum numerical aperture for a given magnification and potentially reduced color correction compared with short-WD objectives designed for coverslips.

Cover glass correction and correction collars

Objectives intended for use with coverslips are corrected for a specific thickness—often approximately 0.17 mm. Some have correction collars that allow tuning for variations in coverslip thickness or for imaging through slightly thicker windows (for example, certain glass-bottom dishes). When used through plastic bottoms or covers of varying thickness and refractive index, objectives without appropriate correction may exhibit spherical aberration and reduced contrast. Where vessel thickness varies among experiments, a correction-collar objective can offer practical flexibility.

Immersion media choices

Immersion objectives use a medium between the front lens and coverslip to improve optical coupling. Common media include water, oil, and silicone oil, each with characteristic refractive index and viscosity. Upright stands handle immersion well on slides because gravity keeps the medium at the interface; inverted stands can also use immersion but require careful application to avoid drips on upward-facing lenses. In both cases, follow the objective’s specified immersion medium to maintain optical correction and protect coatings. For live-cell work, water or silicone oil immersion may better accommodate temperature variations compared to standard oil on biological samples.

Reflected-light (no-cover) objectives

For opaque specimens under epi-illumination, objectives are typically designed for no cover glass and may have long working distance to clear uneven surfaces. These objectives are compatible with both upright metallurgical frames and inverted metallographs. The stand choice will be driven by sample size and access rather than the objectives alone.

Before finalizing a stand purchase, list the objective types you anticipate using, including magnifications, working distances, and correction requirements. Cross-reference them with your sample geometry and contrast methods to avoid incompatibilities.

Imaging and Cameras: Ports, Field Number, and Relay Optics

Most modern stands, upright and inverted, offer trinocular or dedicated camera ports for digital imaging. The camera interface and relay optics influence field of view, sampling, and convenience.

Camera port placement and switching

Trinocular heads or side ports typically include beam splitters to direct a portion of light to the camera. Some stands offer lever-selectable modes, such as 100% to eyepieces or 100% to camera, or split ratios that enable simultaneous viewing and imaging. On inverted stands, camera ports are often integrated into the central observation module, while uprights may place them on a vertical photo tube or side port.

Field number and field of view

The field number characterizes the eyepiece field of view; typical modern eyepieces have field numbers around the low 20s (e.g., approximately 22 mm). The camera’s field of view depends on the camera sensor size and the relay optics used in the camera port. Matching the relay optics to the sensor can optimize sampling without excessive vignetting. Whether upright or inverted, choose relay magnifications and camera sensors that provide appropriate pixel sampling for your objectives and the details you wish to resolve.

Filter cubes and spectral management

For fluorescence, filter cubes (excitation filter, dichroic mirror, emission filter) reside in the epi-illuminator. Both upright and inverted frames accept cube turrets or sliders for rapid channel changes. Keep in mind that high-intensity excitation requires careful light management to minimize photobleaching and background; a well-baffled epi-path and clean optics matter more than stand type in this respect.

Maintenance, Alignment, and Vibration Considerations

Whichever stand you choose, maintaining clean, well-aligned optics pays large dividends. The maintenance emphasis differs slightly between upright and inverted setups because of orientation and accessory placement.

Cleanliness and protection

  • Upright objectives face down; be vigilant about avoiding contact with wet specimens or excessive mounting medium. Use lens paper and appropriate cleaning agents sparingly and in line with manufacturer guidance when needed.
  • Inverted objectives face up; cover the turret when not in use to reduce dust, and consider splash guards or dish holders that minimize spill risk. Wipe vessel bottoms before imaging to remove condensation or fingerprints that would sit directly in the optical path.
  • Condensers accumulate dust at their exposed aperture; whether above (inverted) or below (upright), keep diaphragms and front lenses clean to maintain even illumination.

Alignment and illumination

Aligning Köhler illumination—centering the condenser, focusing the field diaphragm, and adjusting the aperture diaphragm—applies to both stands. For phase contrast, ensure annuli and phase rings are centered. For DIC, verify polarizer and analyzer orientations and prism insertion. While the steps vary by brand, the goals are universal: even fields, controlled angular illumination, and matched pupil planes for the chosen contrast method.

Vibration and environmental stability

High-magnification imaging benefits from vibration control. Inverted stands paired with environmental enclosures may have fans and tubing near the optical path; placing the setup on an anti-vibration platform and isolating pumps or incubator controllers from the bench reduces mechanical and acoustic coupling. Upright stands in teaching rooms are often on sturdy benches; if you extend to high-N.A. imaging, similar vibration mitigation helps.

How to Choose: A Practical Decision Framework

Use the following framework to map your needs to a stand type. This is a planning tool, not brand-specific advice.

1) Define your primary specimen format

  • Mostly slides and thin sections: Upright likely best. The geometry aligns with standard coverslip correction and common transmitted-light condensers.
  • Mostly vessel-based samples (dishes, plates, flasks, chips): Inverted likely best. The unobstructed top enables handling, and long-WD objectives are readily available.
  • Large, heavy, or awkward parts: Inverted metallurgical stands provide a stable platform. For thin, flat coupons and wafers, upright metallurgical stands work well.

2) List your required contrast methods

  • Brightfield/phase: Both are common on uprights and inverteds; confirm condenser compatibility with your vessels if inverted.
  • DIC: Ensure the frame supports DIC modules and that your objectives and condenser are DIC-capable.
  • Darkfield: Uprights make transmitted darkfield straightforward on slides; inverteds may need specialized condensers or reflected darkfield for opaque specimens.
  • Fluorescence: Both support epi-fluorescence; inverteds integrate well with environmental control for live-cell imaging.

3) Check objective compatibility

  • Coverslip correction: If you rely on #1.5 coverslips, standard objectives are fine; if you image through vessel bottoms, consider LWD or correction-collar options.
  • Immersion strategy: Plan how you will apply and maintain the immersion medium, especially on inverted stands where objectives face up.
  • Reflected-light work: Choose no-cover objectives for metallurgical inspection; both stands support epi-modules.

4) Consider ergonomics and workflow

  • Session length: If you sit for extended periods manipulating vessels, inverted eyepiece height can be comfortable; for slide scanning, upright controls are optimized for rapid X–Y navigation.
  • Accessory clearance: Inverted stands leave the specimen top free for pipettes or probes; uprights often have rich accessory ecosystems for polarizing and specialized stages.

5) Plan for environmental control and stability

  • Live imaging: Inverted stands pair naturally with stage-top incubators and enclosures.
  • Fixed samples: Upright stands excel for consistent, repeatable slide imaging with simpler environmental needs.

6) Budget and expandability

  • Base cost: Inverted frames with transmitted-light and epi modules often cost more than comparable uprights due to mechanical complexity and size. Factor in the cost of appropriate objectives (e.g., LWD, DIC-capable) and condensers.
  • Future modules: Verify that the stand supports the contrast methods you might add later (DIC, epi-fluorescence turrets, motorized stages). Infinity-corrected modular stands facilitate future expansion.

As you finalize a selection, revisit your sample geometry and illumination methods to ensure everything aligns: objectives, condensers, epi-modules, and camera ports should work as a coherent system.

Frequently Asked Questions

Can an inverted microscope achieve the same image quality as an upright on slides?

Yes, when equipped with appropriate objectives and condensers, an inverted stand can produce excellent images of specimens mounted on coverslips, particularly using glass-bottom vessels designed with #1.5 coverslip windows. However, for routine slide scanning and certain high-magnification immersion tasks, an upright’s geometry and condenser options may provide a more straightforward setup. The best choice depends on whether your routine work primarily involves slides or vessels.

Is one stand type better for fluorescence?

Neither is inherently better; both support epi-fluorescence with suitable filter cubes and optics. The difference lies in sample handling and environment. Uprights work well for prepared slides and fixed specimens. Inverteds are often preferred for live-cell fluorescence because environmental enclosures fit easily and the sample can be manipulated from above while imaging from below. The right selection mirrors your specimen format and stability needs, as discussed in Where Each Design Excels.

Final Thoughts on Choosing the Right Upright or Inverted Microscope

Selecting between an upright and an inverted microscope is ultimately about harmonizing specimen format, illumination modes, objective compatibility, and workflow ergonomics. Uprights align naturally with slides, thin sections, and polarized-light applications, while inverted stands are the go-to for vessel-based samples, live-cell work, microfluidics, and heavy parts in reflected light. Both designs can host rich ecosystems of contrast methods—brightfield, phase, DIC, darkfield, and fluorescence—when configured with the right condensers, prisms, and epi-illuminators.

If you are at the crossroads, sketch your primary use cases and map them against the decision framework in this article. Pay particular attention to sample geometry and working distance as these parameters drive objective selection and, by extension, the stand that will serve you best. Aim for a coherent system: objectives matched to coverslip or vessel thickness, condensers aligned for your contrast methods, and camera ports configured for the field of view and sampling you need.

The microscope and its revelations (1901) (14763923971)
Attribution: Internet Archive Book Images.

Identifier: microscopeitsrev00carp (find matches)
Title: The microscope and its revelations
Year: 1901 (1900s)
Authors: Carpenter, William Benjamin, 1813-1885 Dallinger, W. H. (William Henry), 1842-1909
Subjects: Microscopy Microscopes Natural history
Publisher: Philadelphia, P. Blackiston’s Sons and Co.
Contributing Library: MBLWHOI Library
Digitizing Sponsor: MBLWHOI Library

View Book Page: Book Viewer
About This Book: Catalog Entry
View All Images: All Images From Book
Click here to view book online to see this illustration in context in a browseable online version of this book.

Text Appearing Before Image:
ws on to the rod and bruises it. This lamp, ifmade, as it should be, with a japanned tin reservoir and a cast-iron tripod foot, is quite inexpensive. There is no justificationfor a circular foot, except that it can be readily and well finishedin the lathe with better apparent results and less labour than otherforms. A small lamp is made by Messrs. R. and J. Beck. We illus-trate it in. fig. 342. The base, A, consists of a heavy ring, into which a square brass 1 It is very important to remove the metal chimney after use, or at least not toleave it on when not in use, since the evaporating paraffin gathers round it and causesundesirable scent when the lamp is again lit. The thinnest slips should be used. 406 MANIPULATION AND PRESERVATION OF THE MICROSCOPE rod, )>. is screwed. The square rod carries a socket, 0, with an arm,D, to which the lamp is attached. On each side of the burner, and attached to the arm, I), is anupright rod, G, to one of which the chimney is fixed, independent of
Text Appearing After Image:
reservoir of the lamp, thus enabling the observer to revolve theni-r and reservoir, and ol.iaiii cither the edge or the flat side ofthe Maine without altering the posilion of the chimney. Thechimney, I1, is made of thin l»ra». witli t\\ openings opposite toother, into \\hich slide 3×1 glass slips of either white, blue, or LAMP WITH LATERAL MOTION 407 opal glass, the latter serving as a reflector ; but we do not considerthe reflexion here accomplished as other than an error; it causesdouble reflexion and confuses the condensed image. A semicircle swings from the two uprights, C4, to which it isattached by the pins, H, placed level with the middle of the flame ;to this semicircle is fixed a dovetailed bar. L. carrying a slidingfitting, O, which bears a Herschel bulls-eye. P. This is complex,in id therefore costly. The bulls-eye is fixed at any inclination by a milled head workingin a. slotted piece of brass, K,fixed to the arm, D. For use with the micro-scope in an upright pos

Note About Images

Please note that these images are extracted from scanned page images that may have been digitally enhanced for readability – coloration and appearance of these illustrations may not perfectly resemble the original work.

We will continue rotating through microscope types and practical decision criteria in upcoming articles. If you found this guide helpful and want more deep-dives on optical design, accessories, and real-world imaging tips, consider subscribing to our newsletter to receive new posts as soon as they are published.