Kfhler Illumination: Resolution, Contrast, and Setup Guide

Table of Contents

What Is Knullfhler Illumination in Optical Microscopy?

Knullfhler illumination (often written Koehler illumination) is the standard method of brightfield illumination used in light microscopy to achieve even, glare-free lighting and to maximize the resolving power of the objective. Instead of projecting an image of the lamp or LED directly onto the specimen, Knullfhler illumination uses a set of conjugate optical planes to homogenize the light at the sample and to control both the size of the illuminated field and the angular distribution of rays entering the objective. When correctly aligned, it yields uniform brightness across the field of view, high contrast, and predictable resolution.

Köhler Illumination with the Upright Microscope (15177755065)
Ask your ZEISS account manager for a lab poster! You’ll find more knowledge brochures and materials on our website www.zeiss.com/microscopy. Images donated as part of a GLAM collaboration with Carl Zeiss Microscopy – please contact Andy Mabbett for details.
Artist: ZEISS Microscopy from Germany

The concept rests on two separable and adjustable diaphragms: the field diaphragm, which defines the illuminated area in the specimen plane, and the aperture diaphragm (at or near the condenser back focal plane), which sets the effective illumination numerical aperture (NA). Knullfhler illumination decouples these two controls: the field diaphragm affects image framing and stray light, while the aperture diaphragm affects resolution, contrast, depth of field, and diffraction blur.

Because Knullfhler provides incoherent illumination for brightfield imaging, the optical performance is governed by classical incoherent imaging theory. This has practical consequences for how the condenser NA should be matched to the objective NA and for how the illumination affects contrast transfer across spatial frequencies in the specimen. If these ideas feel abstract, the next section on conjugate planes and apertures will make the geometry concrete.

Optical Foundations: Conjugate Planes, Aperture vs Field

To understand why Knullfhler illumination works so well, it helps to visualize the microscope as two coupled but distinct subsystems: one for imaging the specimen onto the detector or eyepiece, and another for shaping the illumination that reaches the specimen. These subsystems intersect at the specimen plane but have their own series of conjugate planes where certain elements are imaged sharply.

Conjugate planes in Knullfhler illumination

There are two stacks of conjugate planes, each with different elements in focus:

  • Field (image) conjugates: field diaphragm null specimen plane null intermediate image(s) null detector/retina.
  • Aperture (pupil) conjugates: light source (lamp or LED emitter) null condenser aperture diaphragm/back focal plane null objective rear pupil null camera entrance pupil.
The microscope and its revelations (1901) (14580238850)
FIG. 30.—Path of a ray of light through a modern combination of lenses for compound microscope.
Artist: Internet Archive Book Images

In proper Knullfhler alignment, the field diaphragm is imaged onto the specimen plane, allowing you to control the illuminated area crisply. Meanwhile, the image of the light source is focused at the condenser aperture diaphragm and the objective pupil, not on the specimen. This means the filament structure or LED die is not imaged into the sample plane, eliminating filament granularity and hotspots.

Because the illumination subsystem is telecentric at the specimen plane (illumination rays span a controlled cone determined by the condenser NA), the sample experiences a spatially uniform field of light with a tunable angular distribution. This separation of field size and ray angle is what lets you independently adjust glare (field diaphragm) and resolution/contrast (aperture diaphragm).

Aperture versus field: two different knobs

The terms “aperture” and “field” sometimes get conflated. They regulate different aspects:

  • Field diaphragm: Limits the illuminated specimen area. Closing it reduces stray light from outside the region of interest and can improve contrast by cutting glare from off-scene regions.
  • Aperture diaphragm: Sets the illumination cone half-angle and therefore the illumination NA. It directly influences resolution, contrast, and depth of field by controlling the range of spatial frequencies that are excited and transferred.

When adjusting Knullfhler, always remember: field for area, aperture for angle. We will return to this in the alignment walkthrough to cement the process.

The Role of the Condenser: Numerical Aperture and Aperture Diaphragm

The condenser is the lens group that focuses illumination light onto the specimen, shaping the angular spread of rays. Its performance is described by its numerical aperture (NA), just like an objective. Numerical aperture is a dimensionless measure of a system’s light-gathering and resolving capability:

NA = n * sin(nullb8)

where n is the refractive index of the medium between the condenser and specimen (typically air, n nullc0 1.0, or immersion oil, n nullc0 1.515) and nullb8 is the half-angle of the widest cone of light that can enter or exit the optical element. Higher NA means the condenser can deliver a broader angular distribution of illumination rays, which is essential for resolving fine detail under incoherent brightfield.

Matching condenser NA to objective NA

Under Knullfhler illumination, the condenser NA should generally match the objective NA for optimal resolution in brightfield. That is, if you are using a 0.65 NA objective, aim to set the condenser aperture diaphragm so that the illumination NA is comparable, not drastically smaller. When the illumination NA is much lower than the objective NA, the system behaves more coherently, and the effective bandwidth of transferred spatial frequencies shrinks, reducing resolution and altering contrast. Conversely, opening the aperture to very high NA can maximize resolution but may reduce depth of field and exaggerate sample-induced glare if the specimen is thick or scattering.

Two widely referenced relationships illustrate the interplay of NA and resolution under incoherent brightfield illumination:

  • Rayleigh-style point resolution (assuming adequate, matched illumination): d nullc0 0.61 * nullbb / NA_obj, where nullbb is the wavelength and NA_obj the objective NA.
  • Incoherent cut-off (frequency) perspective: the system’s spatial-frequency cut-off scales with (NA_obj + NA_cond) / nullbb, giving a feature size on the order of d nullc0 nullbb / (NA_obj + NA_cond). This highlights the role of condenser NA when it is not matched to the objective.

Both views are useful. The first reminds us that, with sufficient illumination NA, the objective primarily sets the limit. The second emphasizes that an underset condenser aperture can throttle performance. In practical brightfield microscopy, opening the condenser aperture to around 70null100% of the objective NA is a common guideline for a balance of resolution and contrast. The exact percentage depends on the specimen and imaging goals.

Aperture diaphragm as an illumination NA control

The aperture diaphragm is typically located at or imaged to the condenser’s back focal plane. Adjusting it changes the illumination NA without changing the illuminated field size at the specimen. Closing it reduces the angular spread of light, improving depth of field and increasing apparent contrast in low-frequency features but lowering resolution and potentially introducing diffraction blur and edge ringing. Opening it increases resolution and brightness but can emphasize out-of-focus haze in thick specimens.

The microscope and its revelations (1901) (14764658044)
Please note that these images are extracted from scanned page images that may have been digitally enhanced for readability; coloration and appearance of these illustrations may not perfectly resemble the original work.
Artist: Carpenter, William Benjamin, 1813-1885; Dallinger, W. H. (William Henry), 1842-1909

For best performance, set the aperture diaphragm deliberately for your target. For maximum lateral resolution of thin, high-contrast specimens, open it to match the objective NA. For relief-like contrast on thicker or low-contrast specimens, close it moderately to trade some resolution for higher depth of field and micro-contrast. The trade-offs are central to Knullfhler.

Resolution, Contrast, and Depth of Field Under Knullfhler

Knullfhler illumination places the specimen at a uniform field with a controllable range of incident angles, which strongly influences three interrelated aspects of image quality: resolution, contrast, and depth of field (DOF). Understanding the physics helps you make principled adjustments rather than chasing a subjective “best-looking” image.

Resolution and the transfer of fine detail

In incoherent brightfield, spatial frequencies in the specimen are transferred to the image with a modulation transfer function (MTF) that depends on both the objective pupil and the condenser pupil. When the condenser NA is sufficiently large and matched to the objective NA, the effective cutoff frequency approaches (NA_obj + NA_cond) / nullbb. In that matched-NA regime, the classical point resolution simplifies to the frequently cited Rayleigh form:

d nullc0 0.61 * nullbb / NA_obj   (for adequately matched, high-NA illumination)

If the condenser aperture is closed too far, fine spatial frequencies are no longer efficiently illuminated, reducing the achievable resolution even if the objective itself has higher NA. As a rule of thumb, underillumination yields “smoother” but less detailed images, while matched-NA Knullfhler exposes fine details at the cost of lower DOF.

Contrast: low, mid, and high spatial frequencies

Contrast in brightfield depends on how variations in specimen transmittance or phase modulate the transmitted light and how the imaging system transfers those modulations. Larger features (low spatial frequency) remain visible even when the aperture is closed; mid-to-high spatial frequencies suffer first as the aperture narrows. Opening the aperture improves high-frequency contrast but can decrease the apparent contrast of coarse features because the image becomes “busier” with resolved fine structure. Knullfhler illumination makes this tuning predictable and reproducible.

Depth of field, aberrations, and coherence

Depth of field is inversely related to NA: higher NA yields shallower DOF. Closing the aperture diaphragm increases DOF and can also reduce the sensitivity to small focus errors and minor aberrations by narrowing the ray cone. However, very small illumination NAs move the system toward partial coherence, which changes contrast mechanisms and can introduce diffraction artifacts (e.g., ringing around edges). Balance is key: set the aperture with intention, based on specimen thickness and the goals of your observation or image analysis.

Wavelength dependence and color channels

Resolution and diffraction scales with wavelength. Blue light (nullbb shorter) can resolve finer details than red light. In polychromatic imaging, the effective resolution reflects the weighted contribution of wavelengths to the detected signal, and chromatic aberrations can degrade performance if optics are not corrected. While Knullfhler itself does not eliminate chromatic effects, it provides the uniform, glare-free illumination required to approach the theoretical limits set by your optics. For quantitative color work, ensure white balance and consider narrowband filters if needed.

Field Diaphragm Control: Uniformity, Glare, and Stray Light

The field diaphragm resides in a plane conjugate to the specimen and to the intermediate image. In Knullfhler alignment, the edge of this diaphragm appears sharply at the specimen plane when the condenser height is set correctly. It is the primary tool for controlling the illuminated region and minimizing stray light from outside your field of view.

Why the field diaphragm matters

  • Uniformity: Defining the illuminated area reduces vignetting and edge darkening from stray off-axis light paths.
  • Stray light control: Closing it to just beyond the visible field cuts flare from bright features outside your region of interest, improving micro-contrast.
  • Thermal and photobleach management: In transmitted brightfield with powerful sources or in fluorescence illumination, limiting the field reduces unnecessary sample exposure.
Leuchtfeldblende 071213
Field iris diaphragm (covered by a glass plate) built into the stand base of a Zeiss transmitted light microscope (laboratory Prof. Gitter)
Artist: QuodScripsiScripsi

The field diaphragm does not change the angular distribution of light, so it does not directly change resolution. Instead, think of it as a “mask” for the illuminated specimen area. Open it only as far as needed to illuminate the entire field of view and a small margin.

Tip: During alignment, first locate the field diaphragm edge sharply in focus, center it, and then open it until it just disappears from the view. That simple habit prevents a host of illumination artifacts and helps keep your flat-field uniform.

Stepwise Alignment: Educational Walkthrough for Brightfield

While Knullfhler illumination can feel abstract, setting it up follows a repeatable sequence. The following educational walkthrough is generalized to common transmitted-light microscopes with a collector lens, field diaphragm, condenser with aperture diaphragm, and brightfield objective. It is not a laboratory protocol; rather, it is a conceptual guide for students and hobbyists to understand each control’s function.

Köhler Illumination with the Inverted Microscope (15174751101)
Ask your ZEISS account manager for a lab poster! You’ll find more knowledge brochures and materials on our website www.zeiss.com/microscopy. Images donated as part of a GLAM collaboration with Carl Zeiss Microscopy – please contact Andy Mabbett for details.
Artist: ZEISS Microscopy from Germany

1) Start with a mid-power objective

Select an objective around 10null10null10null10null10null10null10null10null10x to 20x. Using an intermediate magnification makes the field diaphragm edge easy to see and affords enough working distance to manipulate the condenser safely. Ensure the specimen is roughly in focus.

2) Close the field diaphragm and focus the condenser

  • Close the field diaphragm until you see a polygonal or circular edge in the field of view.
  • Focus the condenser height until the diaphragm edge appears sharp at the specimen plane. This ensures the field diaphragm is conjugate with the specimen.
  • Use the condenser centering screws to center the diaphragm image in the field of view.

This step sets the geometric alignment. A focused, centered diaphragm edge confirms that the condenser is at the correct height and centered on the optical axis.

3) Open the field diaphragm to just beyond the field of view

Once centered and sharp, open the field diaphragm until it just disappears from the view. This ensures the entire field is illuminated with minimal stray light. You will revisit this setting when you change objectives, as the field of view changes with magnification.

4) Adjust the aperture diaphragm to set illumination NA

Now adjust the aperture diaphragm to tune the illumination NA. If your microscope includes an indicator, set the aperture to roughly match the objective NA (e.g., 0.75 of the objective NA is a common starting point for contrast). In the absence of a scale, you can judge by eye: opening the diaphragm increases brightness and fine detail sharpness but decreases depth of field; closing it does the opposite. You may also use a phase telescope or Bertrand lens to view the back focal plane and match the illumination pupil to the objective pupil more precisely.

5) Verify even illumination and contrast

Scan across the field to verify uniform brightness. If you notice gradients, revisit condenser centering and the field diaphragm position. Fine-tune the aperture diaphragm for your specific specimen. The right setting depends on your goals: documentation of fine structure typically benefits from higher illumination NA; didactic demonstrations of form may tolerate lower NA for more pleasing 3D relief.

Repeat these steps whenever you change objectives or reconfigure the microscope. With practice, the sequence takes less than a minute.

Comparing Knullfhler to Critical Illumination and Other Schemes

Knullfhler is often contrasted with critical illumination, an older method in which the light source is directly imaged into the specimen plane. This comparison highlights Knullfhlernulls strengths and clarifies when alternate schemes might be appropriate.

Knullfhler vs. critical illumination

  • Uniformity: Knullfhler produces a spatially uniform field because the source is not imaged onto the specimen. Critical illumination can imprint source structure (e.g., filament patterns or LED chip boundaries) onto the image, causing hotspots.
  • Glare control: Knullfhler gives independent control of field size (field diaphragm) and angular content (aperture diaphragm). Critical illumination lacks this clean separation, making stray light harder to manage.
  • Resolution predictability: Knullfhler establishes incoherent illumination conditions that match standard brightfield theory. Critical illumination can produce partial coherence effects that complicate contrast.

Other illumination strategies

  • Oblique illumination: Intentionally offsetting the condenser aperture to emphasize edges or create pseudo-relief. While useful for contrast enhancement, it departs from symmetric Knullfhler and trades isotropic resolution for directional contrast.
  • Darkfield: Uses a condenser that prevents direct axial rays from entering the objective; only scattered light from the specimen is imaged. Alignment is more stringent, and objective NA must be lower than or matched to the inner NA of the darkfield condenser stop. Knullfhler principles (field control) still help manage stray light, even though the angular distribution is deliberately hollow.
  • Phase contrast and DIC: Require specialized illumination pupils (phase annuli or shear prisms) that are positioned in the condenser aperture plane. Knullfhler foundations apply; see modality adaptations.
  • Epi-illumination (reflected light): For opaque specimens under metallurgical objectives, the illumination passes through the objective. Knullfhler-like pupil and field control are implemented in the epi-illuminator. The same logic of conjugate field and aperture planes applies.

In sum, Knullfhler is the general-purpose foundation for transmitted brightfield and the conceptual baseline for many other modes. Even when you deviate (e.g., for oblique or darkfield), understanding Knullfhler helps you predict the consequences.

Adapting Knullfhler Illumination for Phase Contrast, DIC, and Fluorescence

Knullfhler is not only for brightfield. The same separation of aperture and field, and the same conjugate planes, underlie several contrast methods. Adapting Knullfhler to each modality centers on controlling what appears in the condenser aperture plane (or its epi-illumination analog) and how the field is delimited.

Phase contrast

Phase contrast converts phase variations (unseen in plain brightfield) into intensity differences by introducing a phase ring in the objective and a matching annular aperture in the condenser. Proper alignment requires the annulus in the condenser aperture plane to be concentric with the objectivenulls phase ring when viewed in the back focal plane (via a phase telescope or Bertrand lens). The field diaphragm should still be set as in Knullfhler: close, center, focus, and open to just outside the field of view. The aperture setting is dictated by the phase annulus geometry, so the standard Knullfhler aperture diaphragm may be replaced by or integrated with the annulus carrier.

Differential Interference Contrast (DIC)

DIC uses a pair of shear prisms (Wollaston or Nomarski) in the condenser and objective to split and recombine beams that are laterally sheared at the specimen. As in Knullfhler, field control remains crucial for stray light suppression. The illumination pupil is modified by the DIC prismnulls shear, but the principle of setting condenser height for a sharp field diaphragm image at the specimen still applies. Aperture adjustments influence contrast and resolution similarly: opening NA increases fine-detail transfer but reduces DOF and may intensify background gradient if bias retardation is set away from extinction.

Fluorescence microscopy (epi-illumination)

In widefield epi-fluorescence, excitation light is delivered through the objective and fluorescence is collected by the same objective. Knullfhler-like alignment occurs in the epi-illuminator, where the field stop and excitation aperture (often the field and condenser planes of the epi-path) must be aligned. Uniform excitation is essential for quantitative imaging, and flat-field correction depends on stable Knullfhler-like illumination in the epi path. The general guidance mirrors transmitted Knullfhler: limit the field to the cameranulls field of view, center the excitation field stop, and adjust the excitation aperture (if present) to balance signal, photobleaching, and resolution in concert with the objective NA.

Reflected-light (metallurgical) brightfield

For opaque samples (metals, semiconductors), the light reflects from the specimen surface. The epi-illuminator includes its own field and aperture planes, often controlled by sliders or iris diaphragms. The same Knullfhler logic applies: define the illuminated area on the sample and tune the angular spread of incidence through the objective pupil. Surface roughness and gloss strongly affect glare and contrast, making the field diaphragm particularly important to reduce haloing from off-scene reflections.

Digital Imaging Considerations: Cameras, Flat-Fielding, and Photometry

Even, predictable illumination matters even more in digital microscopy than in eyepiece viewing. Camera sensors are unforgiving: small gradients, vignetting, and dust shadows that the eye may ignore can confound analysis or degrade aesthetics. Knullfhler illumination, with its uniform field and controllable pupil, is the starting point for quality imaging.

Exposure, dynamic range, and bit depth

Set illumination intensity and exposure so that highlights (bright background in brightfield) do not clip the sensor and shadows retain structure. Knullfhler helps by providing a stable baseline for brightness across the field, making histogram-based exposure more reliable. Adjusting the aperture diaphragm affects image brightness and contrast; compensate exposure accordingly. For quantitative work, maintain consistent aperture settings across images of the same series to preserve comparable transfer characteristics.

Flat-field correction (shade correction)

Even under good Knullfhler alignment, residual nonuniformities arise from optics, camera cover glass, and sensor microlenses. Flat-field correction removes multiplicative shading by dividing by an image of a uniform field. The quality of this correction depends on stable, centered field and aperture settings. Before collecting a flat, ensure Knullfhler is aligned (see walkthrough), open the field just beyond the field of view, and avoid overfilling the objective pupil excessively. Capture the flat at the same focus, objective, aperture, and illumination intensity as your specimen images.

Vignetting and telecentricity

Numerical vignetting occurs when the illumination or imaging pupil clips off-axis rays. Knullfhler alignment minimizes this by matching pupils and centering the condenser. Mechanical vignetting (field edges cut off) results from misaligned field stops or camera optics; center and size the field diaphragm properly, and match the camera relay optics to the microscopenulls intermediate image. In many microscope camera ports, a projection lens is designed to render the system reasonably telecentric, which stabilizes magnification across focus and reduces shading with sensor tilt. Precise Knullfhler in the illumination path complements these designs.

Color fidelity

If you capture color images, the source spectrum and filters interact with the sensornulls Bayer pattern. Knullfhler ensures uniform excitation but not spectral neutrality. For accurate color, establish white balance with a neutral reference or use calibrated illumination. Avoid changing aperture settings mid-session if color comparability matters; stopping down alters relative spectral weighting through angle-dependent filter responses in some systems.

Diagnosing Artifacts: Dust Shadows, Hotspots, and Uneven Illumination

Even careful users encounter artifacts. The geometry of Knullfhler provides clues to diagnose them. Because the field and aperture planes are distinct, the location of a contaminant in the optical train determines whether it appears as a sharp spot, a diffuse shadow, or vanishes with focus changes.

Recognizing field-plane versus pupil-plane debris

  • Debris in field conjugates (specimen slide, coverslip, camera sensor): Appears sharp or with clear edges when in focus and moves with specimen or camera.
  • Debris in pupil conjugates (condenser aperture, objective pupil): Produces diffuse, out-of-focus shadows or mild shading that changes with aperture setting. Dust at the condenser aperture can cause doughnut-like features when viewed at the back focal plane but is usually invisible in the image plane if Knullfhler is aligned.
  • Debris on the field diaphragm: Typically out of focus in the image unless condenser height is mis-set. If the field diaphragmnulls edge is sharp and correctly positioned, its internal smudges rarely imprint on the image.

Uneven illumination across the field

Brightness falloff can result from several causes:

  • Miscentered condenser: The field diaphragmnulls image is closer to one edge; re-center with condenser screws.
  • Off-axis or defocused collector lens: The lamp or LED emitter is not centered on the optical axis or not properly imaged into the condenser pupil. Realign the collector and source.
  • Incompatible field stop size: The field diaphragm is either too closed (cutting the field) or too open for the intermediate image circle of your objective/camera port.
  • Camera relay mismatch: Sensor sees only part of the intermediate image or experiences optical vignetting in the camera adapter. Adjust or change the relay optics to match the field number.

Hotspots and source structure

If you observe hotspots that resemble the LED die or a lamp filament imprint, the system is not in Knullfhler: the source is being imaged into the specimen plane. Verify the collector lens position and ensure the condenser aperture is in the source conjugate plane. Revisit conjugate planes and the walkthrough.

Back focal plane inspection

A phase telescope or Bertrand lens lets you view the objectivenulls back focal plane directly. Under ideal Knullfhler brightfield, you should see a uniformly illuminated pupil with a circular edge defined by the aperture diaphragm. In phase contrast, you will see the bright annulus aligned with the phase ring. This view is invaluable for diagnosing decentering, asymmetric apertures, or dust in pupil planes.

Frequently Asked Questions

How open should the aperture diaphragm be for best detail?

For thin, high-contrast specimens and when the goal is to resolve the finest detail, open the aperture diaphragm so that the condenser NA roughly matches the objective NA. This ensures high-frequency contrast is transferred efficiently. For example, with a 0.75 NA objective, start near 0.5null0.75 illumination NA and fine-tune by eye. If the specimen is thicker or low-contrast, slightly closing the aperture (to perhaps around 60null70% of the objective NA) can raise apparent contrast and depth of field at the expense of the very finest detail. The optimal setting depends on what information you value in the image.

Do I need to redo Knullfhler every time I change magnification?

Yes, you should at least verify alignment whenever you switch objectives, because the field of view and numerical aperture change. In practice, recheck the field diaphragm edge (close, center, focus, open) and adjust the aperture diaphragm to suit the new objective NA. With experience, this takes less than a minute and pays dividends in uniformity and resolution. If your microscope has parfocal, parcentric objectives and a stable illumination path, small tweaks may be all thatnulls required.

Final Thoughts on Mastering Knullfhler Illumination

Knullfhler illumination is more than a setup step; it is the conceptual foundation that unifies how we think about light in the microscope. By separating control of the illuminated area from the control of the illumination angle, Knullfhler makes image quality tunable and predictable. It provides the uniformity needed for accurate observation and analysis, sets the stage for the objective to reach its resolving limit, and creates a stable baseline for contrast methods such as phase contrast, DIC, and fluorescence.

August Köhler (1866-1948) (8527804902)
August Köhler (1866–1948) was a German professor and early staff member of Carl Zeiss in Jena, Germany, best known for developing the microscopy technique of Köhler illumination, which optimizes resolution by evenly illuminating the field of view.
Artist: ZEISS Microscopy from Germany

The practical takeaways are straightforward:

  • Use the field diaphragm to frame the illuminated region and suppress stray light. Keep it just outside the visible field.
  • Use the aperture diaphragm to set illumination NA thoughtfully: open for fine detail and thin specimens; close slightly for thicker samples or when you need more depth of field.
  • Verify conjugate alignment whenever you change objectives: focus and center the field diaphragm, then tune the aperture.
  • For digital imaging and quantitative work, maintain consistent settings and consider flat-field correction to remove residual shading.

Whether you are a student encountering brightfield for the first time, an educator preparing demonstrations, or a hobbyist pushing your optics to their limits, mastering Knullfhler illumination will transform the clarity and consistency of your images. If you found this deep dive helpful, consider subscribing to our newsletter to receive future articles on microscope fundamentals, contrast methods, and practical optics delivered to your inbox. Explore related topics such as numerical aperture control and digital flat-fielding to continue building your microscopy toolkit.